Inspection System of Vital Sign Monitor
Medical Equipment Center of Sawanpracharak
Equipment Information
IPM NO: 28514 Department : อุบัติเหตุฉุกเฉิน-ER
ID Code: MON-ER-08 IPM Round : รอบที่ 1
IPM date : 7 เม.ย 2569 Mindray / ePM10 / SN: AC708011267
IPM Type : IPM Planning   New-Equipment   Post Repair   IPM Missing Planning   Other
IPM Environment : Ambient Temperature : 24 °C Humidity : 60 %
Equipment Tester Standard
ไม่มีข้อมูล
Equipment Tester Electrical
ไม่มีข้อมูล
This certificate may not be reproduced except in full unless permission for reproduction has been obtained in writing from the calibration.
CONDITION OF THIS RESULT OF TEST

1. THIS CERTIFICATION IS TRACEABLE TO

     -

2. THIS RESULT OF TEST WAS FOUND ACCURATE AS SHOW ON DATE AND PLACE OF TEST ONLY

     -

CheckList

List Set/Indicated Critiria Measured Pass/Fail
1. Leakage Current 500 < 500 μA - N/A
2. Ground Resistance 0.5 ≤ 0.5 Ω - N/A
3. Heart Rate Accuracy 60 bpm 60 ± 5 % bpm 60
4. Heart Rate Accuracy 120 bpm 120 ± 5 % bpm 120
5. Spo2 Rate 90 (% of Spo2) 90 ± 3 % of Spo2 89
6. Spo2 Rate 96 (% of Spo2) 96 ± 3 % of Spo2 95
7. Spo2 Rate 100 (% of Spo2) 100 ± 3 % of Spo2 99
8. Systolic 80 mmHg 80 ± 10 bpm 81
9. Diastolic 50 mmHg 50 ± 10 bpm 52
10. Systolic 120 mmHg 120 ± 10 bpm 117
11. Diastolic 80 mmHg 80 ± 10 bpm 78
12. Systolic 150 mmHg 150 ± 10 bpm 149
13. Diastolic 100 mmHg 100 ± 10 bpm 95
14. Alarm Operation Device pass
Note

Acceptor BY

..............................

หัวหน้าหน่วยงาน

Done BY

..............................

(ประจักษ์ ชัยศิริ)