Inspection System of Electrocardiographs
Medical Equipment Center of Sawanpracharak
Equipment Information
IPM NO: 28341 Department : ศัลยกรรมเฉพาะทาง
ID Code: ECG-SurgSP-01 IPM Round : รอบที่ 1
IPM date : 16 ม.ค 2569 GE Healthcare / MAC5 / SN: SUE22440012WA
IPM Type : IPM Planning   New-Equipment   Post Repair   IPM Missing Planning   Other
IPM Environment : Ambient Temperature : 24 °C Humidity : 60 %
Equipment Tester Standard
MASTER : Vital Sign Simulaotr
FLUKE / Prosim 8 / SN: 5065015
Cal Date Lasted : 21 ก.ค 2567
Equipment Tester Electrical
MASTER : Electrical Safety Analyzer
FLUKE / ESA620 / SN: 5653005
Cal Date Lasted : 18 ก.ค 2567
This certificate may not be reproduced except in full unless permission for reproduction has been obtained in writing from the calibration.
CONDITION OF THIS RESULT OF TEST

1. THIS CERTIFICATION IS TRACEABLE TO

     -

2. THIS RESULT OF TEST WAS FOUND ACCURATE AS SHOW ON DATE AND PLACE OF TEST ONLY

     -

CheckList

List Set/Indicated Critiria Measured Pass/Fail
1. Ground Resistance ≤ 0.5 Ω 0.5 ≤ 0.5 Ω - N/A
2. Chassis Leakage Current ≤ 500 μA 500 ≤ 500 μA - N/A
3. Petient Leakage Current (Grounded) ≤ 100 μA 100 ≤ 100 μA - N/A
4. Petient Leakage Current (UnGrounded) ≤ 500 μA 500 ≤ 500 μA - N/A
5. Heart Rate Accuracy 60 bpm 60 ± 5 % 60
6. Heart Rate Accuracy 120 bpm 120 ± 5 % 120
7. Paper Speed 25 mm/sec 25 ± 2 % 25
8. Paper Speed 50 mm/sec 50 ± 2 % 50
Note

Acceptor BY

..............................

หัวหน้าหน่วยงาน

Done BY

..............................

(กัมพล พุทธเจริญ)