Inspection System of INFUSION PUMP
Medical Equipment Center of Sawanpracharak
Equipment Information
IPM NO: 28183 Department : สงฆ์อาพาธ
ID Code: INP-PU-20 IPM Round : รอบที่ 1
IPM date : 26 ธ.ค 2568 Fresinius Kabi / Agilia MC / SN: 22805252
IPM Type : IPM Planning   New-Equipment   Post Repair   IPM Missing Planning   Other
IPM Environment : Ambient Temperature : 24 °C Humidity : 60 %
Equipment Tester Standard
MASTER : Infusion Device Analyzer
FLUKE / IDA-5 / SN: 5640072
Cal Date Lasted : 18 ก.ค 2567
Equipment Tester Electrical
MASTER : Electrical Safety Analyzer
FLUKE / ESA620 / SN: 5653005
Cal Date Lasted : 18 ก.ค 2567
This certificate may not be reproduced except in full unless permission for reproduction has been obtained in writing from the calibration.
CONDITION OF THIS RESULT OF TEST

1. THIS CERTIFICATION IS TRACEABLE TO

     -

2. THIS RESULT OF TEST WAS FOUND ACCURATE AS SHOW ON DATE AND PLACE OF TEST ONLY

     -

CheckList

List Set/Indicated Critiria Measured Pass/Fail
1. Leakage Current 500 < 500 μA - N/A
2. Ground Resistance 0.5 ≤ 0.5 Ω - N/A
3. Flow Rate 10 ml/hr 10 ± 5 % 10.23
4. Flow Rate 100 ml/hr 100 ± 5 % 100.74
5. Flow Rate 300 ml/hr 300 ± 5 % 301.54
6. Occlusion < 850mmHg 850 < 850 mmHg 447
7. Air In Line Detection 100 < 100 μL - N/A
Note

Acceptor BY

..............................

หัวหน้าหน่วยงาน

Done BY

..............................

(กัมพล พุทธเจริญ)