Inspection System of SYRINGE PUMP
Medical Equipment Center of Sawanpracharak
Equipment Information
IPM NO: 28176 Department : ศูนย์เครื่องมือแพทย์
ID Code: SRP-FS-MEC-01 IPM Round : รอบที่ 1
IPM date : 26 ธ.ค 2568 Fresinius Kabi / Agilia / SN: 23255417
IPM Type : IPM Planning   New-Equipment   Post Repair   IPM Missing Planning   Other
IPM Environment : Ambient Temperature : 24 °C Humidity : 60 %
Equipment Tester Standard
ไม่มีข้อมูล
Equipment Tester Electrical
ไม่มีข้อมูล
This certificate may not be reproduced except in full unless permission for reproduction has been obtained in writing from the calibration.
CONDITION OF THIS RESULT OF TEST

1. THIS CERTIFICATION IS TRACEABLE TO

     -

2. THIS RESULT OF TEST WAS FOUND ACCURATE AS SHOW ON DATE AND PLACE OF TEST ONLY

     -

CheckList

List Set/Indicated Critiria Measured Pass/Fail
1. Leakage Current 500 < 500 μA - N/A
2. Ground Resistance 0.5 ≤ 0.5 Ω - N/A
3. Flow Rate 10 ml/hr 10 ± 5 % 9.33
4. Flow Rate 50 ml/hr 50 ± 5 % - N/A
5. Flow Rate 150 ml/hr 150 ± 5 % - N/A
6. Limit total Rate 10 ml 10 ± 3 % - N/A
7. Limit total Rate 20 ml 20 ± 3 % - N/A
8. Limit total Rate 50 ml 50 ± 3 % - N/A
9. Occlusion < 850mmHg 850 < 850 mmHg - N/A
10. Air In Line Detection 100 < 100 μL - N/A
Note

Acceptor BY

..............................

หัวหน้าหน่วยงาน

Done BY

..............................

(กัมพล พุทธเจริญ)