Inspection System of Defibrillator
Medical Equipment Center of Sawanpracharak
Equipment Information
IPM NO: 17770 Department : อายุรกรรมหญิง1
ID Code: DEF-MedF1-04 IPM Round : รอบที่ 2
IPM date : 14 พ.ค 2568 Zoll / R Series / SN: AF21F137559
IPM Type : IPM Planning   New-Equipment   Post Repair   IPM Missing Planning   Other
IPM Environment : Ambient Temperature : 24 °C Humidity : 60 %
Equipment Tester Standard
MASTER : Defibrillator Analyzer
FLUKE / IMPULSE 7000DP / SN: 5067010
Cal Date Lasted : 18 ก.ค 2567
Equipment Tester Electrical
MASTER : Electrical Safety Analyzer
FLUKE / ESA620 / SN: 5653005
Cal Date Lasted : 18 ก.ค 2567
This certificate may not be reproduced except in full unless permission for reproduction has been obtained in writing from the calibration.
CONDITION OF THIS RESULT OF TEST

1. THIS CERTIFICATION IS TRACEABLE TO

     -

2. THIS RESULT OF TEST WAS FOUND ACCURATE AS SHOW ON DATE AND PLACE OF TEST ONLY

     -

CheckList

List Set/Indicated Critiria Measured Pass/Fail
1. Leakage Current 500 < 500 μA - N/A
2. Ground Resistance 0.5 ≤ 0.5 Ω - N/A
3. Output Energy 50 jules 50 ± 15 % 54
4. Output Energy 100 jules 100 ± 15 % 109.2
5. Output Energy 200 jules 200 ± 15 % 231.7
6. Synchronized Operation 60 < 60 mm/sec 15
7. Charge Time And Discharge 10 Times 10 < 10 sec 5.6
8. Internal Paddle Energy 50 jules 50 < 50 jules - N/A
9. AED Mode Analysis pass
10. Heart Rate Accuracy 60 bpm 60 ± 5 % bpm 60
11. Heart Rate Accuracy 120 bpm 120 ± 5 % bpm 120
12. SpO2 Rate 90 (% of SpO2) 90 ± 3 % of SpO2 90
13. SpO2 Rate 96 (% of SpO2) 96 ± 3 % of SpO2 96
14. SpO2 Rate 100 (% of SpO2) 100 ± 3 % of SpO2 100
15. Systolic 80 mmHg 80 ± 10 mmHg 83
16. Diastolic 50 mmHg 50 ± 10 mmHg 51
17. Systolic 120 mmHg 120 ± 10 mmHg 125
18. Diastolic 80 mmHg 80 ± 10 mmHg 82
19. Systolic 150 mmHg 150 ± 10 mmHg 153
20. Diastolic 100 mmHg 100 ± 10 mmHg 99
21. Alarm Operation Device pass
Note

Acceptor BY

..............................

หัวหน้าหน่วยงาน

Done BY

..............................

(ประจักษ์ ชัยศิริ)